R&S®TS-EMF EMF Measurement System
Easy, frequency-selective measurement of EMF emissions
Specifications
| Isotropic antennas |
|
|
|
| Measurement principle | isotropic reception due to orthogonally arranged antenna elements that are electronically switched | ||
| Frequency range | 9 kHz to 200 MHz | 30 MHz to 3 GHz | 700 MHz to 6 GHz |
| Field-strength measurement range | approx. 1 mV/m to 100 V/m | ||
| Isotropy | ≤ ±1.37 dB | ≤ ±2.1 dB ±1.0 dB (f = 900 MHz) ±1.7 dB (f = 1800 MHz) |
≤ ±2.5 dB (0.6 GHz to 2 GHz) ≤ ±2.2 dB (2 GHz to 3.6 GHz) ≤ ±3.0 dB (3.6 GHz to 6 GHz) |
| Mechanical design | radome protection against mechanical damage and environmental hazards | ||
| Antenna factor | individual calibration data, saved on USB dongle and/or CD | ||
| Axis switching | RF solid state switch | ||
| Connecting cables | direct connection to spectrum analyzer (extension cable length 8 m, see accessories) |
integrated cable length 2 m, ferritebeaded (extension cable length 8 m, see accessories) | integrated cable length 2 m |
| RF connector | N male | ||
| Connector, control line | 7-pin connector (binder) for direct connection to R&S®FSH power sensor port, adapter cable for connection to notebook or spectrum analyzer with USB output | ||
| Tripod adapter | ¼-inch thread, quick connector for antenna | ||
| Ambient conditions | –10 °C to +50 °C, safety class IP54 (details see user manual) | ||
| Weight (incl. cable) | 0.85 kg | 1.3 kg | 0.95 kg |
| Dimensions | 550 mm, Ø 146 mm |
475 mm, Ø 170 mm |
415 mm, Ø 87 mm |
| Power supply | via spectrum analyzer or laptop | ||
| Requirements for notebook (not supplied) |
MS WindowsXP, XP Tablet PC Edition, Vista operating system, free hard disk space min. 10 Mbytedisplay resolution min. 800 × 600, two USB interfaces interface to test instrument (depending on analyzer: GPIB, LAN, USB or FireWire) recommended application: MS Excel 2000, XP, 2007 |
||
| Equipment supplied | isotropic sensor with connecting cable Operator’s guide for |
||
| Expanded measurement uncertainty (95 % confidence level (k = 2)) |
≤ ±2.5 dB | ≤ ±3.1 dB ±2.3 dB at 0.9 GHz ±2.8 dB at 1.8 GHz |
≤ ±3.4 dB (0.6 GHz to 2 GHz) ≤ ±3.2 dB at (2 GHz to 3.6 GHz) ≤ ±3.9 dB at (3.6 GHz to 6 GHz) |
| Expanded measurement uncertainty (95 % confidence level (k = 2)) |
≤ ±2.4 dB | ≤ ±3.0 dB ±2.2 dB at 0.9 GHz ±2.7 dB at 1.8 GHz |
≤ ±3.2 dB (0.6 GHz to 2 GHz) ≤ ±2.9 dB at (2 GHz to 3.6 GHz) ≤ ±3.8 dB at (3.6 GHz to 6 GHz) |