Specifications

Isotropic antennas R&S®TSEMF-B3 R&S®TSEMF-B1 R&S®TSEMF-B2
Measurement principle isotropic reception due to orthogonally arranged antenna elements that are electronically switched
Frequency range 9 kHz to 200 MHz 30 MHz to 3 GHz 700 MHz to 6 GHz
Field-strength measurement range approx. 1 mV/m to 100 V/m
Isotropy ≤ ±1.37 dB ≤ ±2.1 dB
±1.0 dB (f = 900 MHz)
±1.7 dB (f = 1800 MHz)
≤ ±2.5 dB (0.6 GHz to 2 GHz)
≤ ±2.2 dB (2 GHz to 3.6 GHz)
≤ ±3.0 dB (3.6 GHz to 6 GHz)
Mechanical design radome protection against mechanical damage and environmental hazards
Antenna factor individual calibration data, saved on USB dongle and/or CD
Axis switching RF solid state switch
Connecting cables direct connection to spectrum analyzer
(extension cable
length 8 m, see accessories)
integrated cable length 2 m, ferritebeaded (extension cable length 8 m, see accessories) integrated cable length 2 m
RF connector N male
Connector, control line 7-pin connector (binder) for direct connection to R&S®FSH power sensor port, adapter cable for connection to notebook or spectrum analyzer with USB output
Tripod adapter ¼-inch thread, quick connector for antenna
Ambient conditions –10 °C to +50 °C, safety class IP54 (details see user manual)
Weight (incl. cable) 0.85 kg 1.3 kg 0.95 kg
Dimensions 550 mm,
Ø 146 mm
475 mm,
Ø 170 mm
415 mm,
Ø 87 mm
Power supply via spectrum analyzer or laptop
Requirements for notebook
(not supplied)
MS WindowsXP, XP Tablet PC Edition, Vista operating system, free hard disk space min. 10 Mbytedisplay resolution min. 800 × 600, two USB interfaces
interface to test instrument (depending on analyzer: GPIB, LAN, USB or FireWire) recommended application: MS Excel 2000, XP, 2007
Equipment supplied isotropic sensor with connecting cable
Operator’s guide for R&S®EMF R&S®RFEX and R&S®RFEX-Fast measurement software ( on CD), tripod adapter,USB adapter cable, transit case
Expanded measurement uncertaintyR&S®TS-EMF with R&S®FSH
(95 % confidence level
(k = 2))
≤ ±2.5 dB ≤ ±3.1 dB
±2.3 dB at 0.9 GHz
±2.8 dB at 1.8 GHz
≤ ±3.4 dB (0.6 GHz to 2 GHz)
≤ ±3.2 dB at (2 GHz to 3.6 GHz)
≤ ±3.9 dB at (3.6 GHz to 6 GHz)
Expanded measurement uncertaintyR&S®TS-EMF with R&S®FSL
(95 % confidence level
(k = 2))
≤ ±2.4 dB ≤ ±3.0 dB
±2.2 dB at 0.9 GHz
±2.7 dB at 1.8 GHz
≤ ±3.2 dB (0.6 GHz to 2 GHz)
≤ ±2.9 dB at (2 GHz to 3.6 GHz)